<1>, When taking the instrument out of the packaging box, please check whether the accessories and warranty card are complete.
<2>, Check the instrument appearance and interior for any serious damage caused by transportation or other abnormal phenomena.
<3>, Preparation before startup and connection:
(1) Use the "filter" from the accessories to replace the sealing plug on the "Air Inlet" port on the back panel of the instrument.
(2) Tighten the sealing cap on the "Air Outlet" port on the back panel to ensure no air leakage.
(3) Connect the grounding terminal (middle piece) of the instrument power cord and the chassis shell reliably to the ground.
(4) Connect the power supply, turn on the switch, and the instrument starts working. At this time, the output pressure gauge pointer gradually rises to 0.4 Mpa (set at the factory). After a few minutes, the compressor stops working (set internally). At this time, loosen the sealing nut on the air output A slightly to allow some exhaust. Soon, the compressor will automatically restart. During the repeated cycle of compressor start-stop-start, the pressure gauge should always remain at 0.4Mpa, indicating that the instrument is normal. (When the exhaust volume exceeds the rated value, the pressure gauge will not stabilize at 0.4Mpa).
(5) During the start-stop cycle of the compressor, each time it stops working, it will automatically discharge some water, but this cannot completely replace manual drainage, so frequent manual drainage is still required.
(6) After the above is normal, turn off the power, remove the pressure cap on the air output port, connect the pipeline with the appropriate outer diameter to the instrument, and ensure a sealed and leak-free connection.
(7) For subsequent operation, simply turn on the power switch.
(8) If only one output is used, seal the other outlet with a blind cap.
Precautions:
1. Since the compressor is an inductive load, the instantaneous current during power on/off is several times higher than normal operation, making it easier to blow fuses. An 8A fuse should be used.
2. If the compressor does not start during operation and the thermal protection relay activates, it indicates that the compressor temperature is too high. It will automatically recover after cooling down.
3. To avoid excessive water accumulation inside the machine affecting air purity, press the drain switch for a few seconds before each shutdown. (You can connect the drain outlet to a plastic bottle to prevent water splashing). Drain at least once a day.
4. Observe the color-changing silica gel through the observation window. Users can replace the silica gel with new one or perform drying treatment as needed.
5. To ensure gas purity, the activated carbon needs to be replaced every 1000 hours of operation (activated carbon is 20-40 mesh, pencil lead type).
6. Since the compressor operates in an open circuit mode, lubricating oil will be discharged with water, causing consumption. Therefore, after one year of use, appropriately add lubricating oil to the compressor. The oil filling port is next to the compressor outlet (or can be added from the inlet). It is recommended to add 200 grams of No. 18 refrigeration oil.
7. The inlet filter needs regular cleaning (the cycle depends on indoor dust conditions; ultrasonic cleaning can be used) to maintain smooth air intake. Otherwise, it may cause increased load and heating of the compressor, and if the temperature is too high, overheating protection may occur, leading to shutdown.
Purpose and Functions:
Intelligent Dry Air Generator, specially designed to provide high-purity dry air with a dew point as low as -60℃~-70℃ for large power equipment such as transformers and reactors during installation and maintenance. It can replace traditional drying methods such as hot oil circulation, nitrogen supplementation, and vacuum drying, and is more efficient, economical, safe, and environmentally friendly.
Intelligent Dry Air Generator, during installation and maintenance of power equipment, is not affected by changes in external climate conditions. Especially for overnight operations, simply seal the flanges and fill with dry air to 0.02-0.03Mpa. There is no need for oil injection, nitrogen supplementation, or vacuum pumping. Transformers, reactors, and other power equipment will not be affected by moisture, greatly reducing workload, shortening construction periods, and improving installation and maintenance quality.
It is also suitable for air drying and circulating drying in warehouses, packaging, electronics, pharmaceuticals, and other industries. It is also applicable for drying nitrogen, generator set hydrogen, and SF6 equipment.
Replaces traditional drying methods, efficient, economical, safe, and environmentally friendly. It integrates air compression, refrigeration drying, and adsorption into one unit and is freely movable.
Product Principle:
Atmospheric air enters the air storage tank through the air compressor. Most of the moisture is compressed and liquefied, then discharged through the drain valve, completing the first drying of the air. Then it enters the refrigerated dryer, where water vapor condenses into water, completing the second drying. Then it enters the adsorption dryer for the third drying, adsorbing the remaining trace moisture. Finally, it passes through a high-precision air filter and is delivered to the equipment requiring dry gas.
Performance and Features:
It mainly consists of four parts: the air source system, refrigeration drying system, adsorption drying system, and electrical control system, forming a mobile high-purity dry air generation device.
Air source system composition: composed of air dust filter, oil-free air pump (or screw oil-free air compressor), radiator (air-cooled), air-water separator, and filter, outputting clean and oil-free gas.
Refrigeration drying system composition: This device mainly uses imported fully enclosed high-temperature compression refrigeration units, with stable and reliable performance, providing greater cooling capacity. The system configuration includes imported fully enclosed compressor units, air heat exchangers, evaporators, air-water separators, and fine filters. In the refrigeration process, R22 refrigerant first passes through the fully enclosed high-temperature compression refrigeration compressor, cooler, desiccant, solenoid valve, and finally returns to the refrigeration compressor through the evaporator, achieving a low-temperature environment of 0-5℃ inside the evaporator.
Electrical control: The main control electrical parts use Schneider Electric; the entire machine has stable, safe, and reliable electrical performance.